Pin assignment
Pin 1 n/c
Pin 2 RXD
Pin 3 TXD
Pins 4, 6 Pins 4 and 6 are connected internally
Pin 5 SHLD-GND
Pins 7, 8 Pin 7 and Pin 8 are connected internally
Pin 9 n/c
Housing SHLD-GND
USB interface Only connect USB storage media that do not
have their own power supply. Do not connect
any charging cables.
Type USB type A connector; USB 1.1
Supported devices USB flash drive formatted with FAT16 or FAT32
Dräger recommends storage media with FIPS
140-2 compatible encryption and their own
encryption facility.
LAN interface Only for Dräger Remote Service
Only connect devices and/or networks that meet
the requirements of IEC 60950-1/IEC 62368-1
for ungrounded SELV circuits and the require-
ments of IEC 60601-1 (as of the 2nd edition) for
touchable secondary circuits with a maximum
nominal voltage of 24 V DC.
Type RJ45 plug
Transfer speed 10Base-T, IEEE 802.3 Clause 14.
Requires at least a CAT3 cable.
Connection for external workstation light
(optional)
Only for workstation lights approved by Dräger,
see list of accessories.
Distribution piece for central O
2 supply
(optional)
Supply pressure 2.7 to 6.9 kPa x 100
39 to 100 psi
0.27 to 0.69 MPa
Maximum permissible extraction flow rate 40 L/min