Technical data
246 Instructions for use Perseus A500 SW 2.0n
Mains power sockets Observe the maximum current for each power
socket, the total current for all power sockets,
and the total permissible leakage current.
Mains power sockets have no battery back-up
and are independent of the switch-on state of
the device.
Power sockets (version without isolation
transformer)
Power socket type IEC
Power sockets with protective ground 2 (maximum 3.15 A per power socket)
Power sockets without protective ground 2 (maximum 1 A per power socket)
Fuse type Fuses, T3,15AH
250V IEC 60127-2/V and
T1AH
250V IEC 60127-2/V, size 5 mm * 20 mm
Power sockets
(version with isolation transformer)
Power socket type Country-specific
Fuse type Electronic automatic circuit breakers: 4 x 3 A,
slow-blow, for voltages up to 250 V and break-
ing capacity 2000 A.
And 1 x 8 A, slow-blow, for voltages up to 250 V
and breaking capacity 2000 A, in each case
conforming to IEC 60934
Power sockets with protective ground 4 (maximum 3 A per power socket)
Total current Max. 8 A
Main switch for isolation transformer and all
power sockets
Device outputs (continued)