Parts List and Schematics
134
Perm Flow
Total flow past the permeation chamber during an activated
internal span mode. In ml/min
135
Perm Oven Setpoint
Set target temperature for the permeation oven. Default is
50°C
136
Perm Oven Temp
Temperature readout of the Permeation oven. Units in °C
137
Ozone Target
Ozone Target for S10 IZS Ozone generation
138
Conc Adjusted 1
Concentration value in PPM before filtering
139
Conc Adjusted 2
140
Conc Adjusted 3
141
Chopper Speed
Revolutions in the last second
142
Chopper Frequency
In rpm
143
IR Source
The S30 IR source voltage
144
Background (hrs)
The background interval in hours.
0.0 if disabled
0.25 for every 15 minutes
0.30 for every 20 minutes
0.50 for every 30 minutes
1.00 for every hour
24.00 for once a day
145
Cycle Time
In minutes
146
CO Cooler Pot
CO Cooler voltage adjustment POT
147
CO Source Pot
CO Source voltage adjustment POT
148
CO Test Meas. Pot
Diagnostics use only
149
CO Test Ref. Pot
Diagnostics use only
150
O3 Ref Average
S10 Background Average
151
PTF Correction (gas 1)
Pressure Temperature Flow Compensation Factor for first gas
152
PTF Correction (gas 2)
Pressure Temperature Flow Compensation Factor for second
gas in dual gas analysers.
153
Inst. Cell Pressure
Instantaneous cell pressure
154
Manifold Pressure
Manifold Pressure in S40 instruments
155
Cell Press. (gas 1)
Cell Pressure for Gas 1
156
Cell Press. (gas 2)
Cell Pressure for Gas 2
157
Cell Press. (Bgnd)
Cell Pressure when in Background
158
Background
0 = the instrument is measuring a gas sample
1 = the instrument is measuring background air
159
Gas To Measure
S51 only; see Measurement Settings Menu
0 = Measure both gasses
1 = Measure SO2 only
2 = Measure H2S only