Serinus® 10 User Manual 3.2
Page 206
24 = Span Measure Aux2
25 = Background Fill Span
26 = Background Measure Span
27 = Background Purge Span
28 = Electronic Zero Adjust
29 = Instrument Warm Up
30 = Background Adjust Fill
31 = Background Adjust Measure
111
CO Lin. Factor A
CO Linearisation Factor A
112
CO Lin. Factor B
CO Linearisation Factor B
113
CO Lin. Factor C
CO Linearisation Factor C
114
CO Lin. Factor D
CO Linearisation Factor D
115
CO Lin. Factor E
CO Linearisation Factor E
116
Instrument Units
0 = ppm
1 = ppb
2 = ppt
3 = mg/m³
4 = µg/m³
5 = ng/m³
6 = %
117
Backgnd Meas. Time
In seconds.
These parameters can be changed, but only temporarily;
restarting the instrument will restore them to their default
values.
118
Sample Fill Time
119
Sample Measure Time
120
Aux Measure Time
121
Aux Smpl. Fill Time
122
Backgnd Fill Time
123
Zero Fill Time
124
Zero Measure Time
125
Span Fill Time
126
Span Measure Time
127
O3 Gen. Coeff D
O3 Generator Coefficient D
128
Backgnd Pause Time
In seconds
129
Bkgnd Intrleav Fact
130
Cal. Pressure 2
Calibration Pressure for 2nd gas
131
2nd Instrument Gain
Only valid with an S40 variant set to multiple gains. Shows the
second gas gain (NO). Default is 1.0.
132
Background voltage
Units in Volts
133
Perm Rate
Permeation rate of the gas in ng/min