Parts List and Schematics
95
Output Type AO 0
Output Type
1 = Voltage
0 = Current
96
Output Type AO 1
97
Output Type AO 2
98
Anlg Ofst/Rng AO 0
Voltage Offset /Current Range
0 = 0% or 0 - 20mA
1 = 5% or 2 - 20mA
2 = 10% or 4 - 20mA
99
Anlg Ofst/Rng AO 1
100
Anlg Ofst/Rng AO 2
101
F/Scale Volt AO 0
5.0 Volt Calibration value
102
F/Scale Volt AO 1
103
F/Scale Volt AO 2
104
Z Adj Volt AO 0
0.5 Volt Calibration value
105
Z Adj Volt AO 1
106
Z Adj Volt AO 2
107
-Analog Supply
Negative analog supply
108
Digital Outputs
A single byte expressing the most recent state of the digital
outputs
109
Digital Inputs
A single byte expressing the most recent state of the digital
inputs
110
Instrument State
0 = Sample Fill
1 = Sample Measure
2 = Sample Fill Aux
3 = Sample Measure Aux
4 = Sample Fill Aux2
5 = Sample Measure Aux2
6 = Background Fill
7 = Background Measure
8 = Background Purge
9 = Background Fill Aux
10 = Background Measure Aux
11 = Zero Fill
12 = Zero Measure
13 = Zero Fill Aux
14 = Zero Measure Aux
15 = Zero Fill Aux2
16 = Measure Aux2
17 = Background Fill Zero
18 = Background Measure Zero
19 = Span Fill
20 = Span Measure
21 = Span Fill Aux
22 = Span Measure Aux
23 = Span Fill Aux2

Table of Contents