Parts List and Schematics
3 = 6 Hrs
4 = 4 Hrs
5 = 2 Hrs
6 = Disable
Note: this parameter is deprecated; while it can still be read,
writing to it will have no effect. Use parameter 144 instead.
167
Service (COM1) Baud
Serial baud rate
0 = 1200 bps
1 = 2400 bps
2 = 4800 bps
3 = 9600 bps
4 = 14400 bps
5 = 19200 bps
6 = 38400 bps
168
Multidrop (COM2) Baud
169
Service Protocol
0 = EC9800
1 = Bayern-Hessen
2 = Advanced
3 = Modbus
170
Multidrop Protocol
171
AO1 Over Range
The Upper Concentration Range when Over-Ranging is
enabled
172
AO2 Over Range
173
AO3 Over Range
174
AO1 Over-Ranging
0 = Over Ranging Disabled
1 = Over Ranging Enabled
2 = Over Ranging enabled and currently active
175
AO2 Over-Ranging
176
AO3 Over-Ranging
177
Heater Set Point
Cell Heater Set Point units in °C
178
High Volt Adjust Pot
High voltage pot setting
179
PMT Test LED Pot
PMT Test LED intensity controller POT
180
Last Power Failure
Time Stamp of the Last power fail (4 byte time stamp)
Bit 31:26 ---- Year (0 99)
Bit 25:22 ---- Month ( 1 12)
Bit 21:17 ---- Date (1 31)
Bit 16:12 ---- Hour (00 23)
Bit 11:06 ---- Min (00 59)
Bit 05:00 ---- Sec (00 59)
181
Inst Manifold Press.
Manifold Pressure in S40 instruments (instantaneous)
182
Cell Press. (Gas 5)
Cell Pressure for Gas 5 (Nx)
183
Gas 4 Inst.
Calculated gas concentration currently displayed on front
screen E.g. NH3
184
Gas 4 Avg.
Average of the readings (for Gas 4) of the last n minutes where
n is the averaging period E.g. NH3