What causes Ecotech Acoem Serinus 10 sample flow to not equal 0.5 slpm +/-0.05?
- YyoungjoseSep 12, 2025
If your Ecotech Measuring Instruments sample flow is not equal to 0.5 slpm +/-0.05 (External pump model only), check or replace the sample filter, pump, and valves. Also, check or replace the sintered filter and re-calibrate the pressure sensors.