Proline Promass E 500 PROFIBUS PA Commissioning
Endress+Hauser 111
10.6.3 Selecting and setting the medium
The Select medium wizard submenu contains parameters that must be configured in
order to select and set the medium.
Navigation
"Setup" menu → Medium selection
‣
Medium selection
Select medium
→  111
Select gas type
→  111
Reference sound velocity
→  112
Temperature coefficient sound velocity
→  112
Pressure compensation
→  112
Pressure value
→  112
External pressure
→  112
Parameter overview with brief description
Parameter Prerequisite Description Selection / User entry / 
User interface
Select medium – Use this function to select the type of
medium: "Gas" or "Liquid". Select the
"Other" option in exceptional cases in
order to enter the properties of the
medium manually (e.g. for highly
compressive liquids such as sulfuric
acid).
• Liquid
• Gas
Select gas type In the Medium selection submenu, the
Gas option is selected.
Select measured gas type. • Air
• Ammonia NH3
• Argon Ar
• Sulfur hexafluoride SF6
• Oxygen O2
• Ozone O3
• Nitrogen oxide NOx
• Nitrogen N2
• Nitrous oxide N2O
• Methane CH4
• Hydrogen H2
• Helium He
• Hydrogen chloride HCl
• Hydrogen sulfide H2S
• Ethylene C2H4
• Carbon dioxide CO2
• Carbon monoxide CO
• Chlorine Cl2
• Butane C4H10
• Propane C3H8
• Propylene C3H6
• Ethane C2H6
• Others