EasyManua.ls Logo

Endress+Hauser Proline Promass E 500 - Parameter Overview with Brief Description

Endress+Hauser Proline Promass E 500
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass E 500 PROFIBUS PA Commissioning
Endress+Hauser 109
Temperature unit
→  110
Pressure unit
→  110
Parameter overview with brief description
Parameter Description Selection Factory setting
Mass flow unit Select mass flow unit.
Effect
The selected unit applies to:
Output
Low flow cut off
Simulation process variable
Unit choose list Country-specific:
kg/h
lb/min
Mass unit Select mass unit. Unit choose list Country-specific:
kg
lb
Volume flow unit Select volume flow unit.
Effect
The selected unit applies to:
Output
Low flow cut off
Simulation process variable
Unit choose list Country-specific:
l/h
gal/min (us)
Volume unit Select volume unit. Unit choose list Country-specific:
l (DN > 150 (6"): option)
gal (us)
Corrected volume flow unit Select corrected volume flow unit.
Effect
The selected unit applies to:
Corrected volume flow parameter
(→  166)
Unit choose list Country-specific:
Nl/h
Sft³/min
Corrected volume unit Select corrected volume unit. Unit choose list Country-specific:
Nl
Sft³
Density unit Select density unit.
Effect
The selected unit applies to:
Output
Simulation process variable
Density adjustment (Expert menu)
Unit choose list Country-specific:
kg/l
lb/ft³
Reference density unit Select reference density unit. Unit choose list Country-specific
kg/Nl
lb/Sft³
Density 2 unit Select second density unit. Unit choose list Country-specific:
kg/l
lb/ft³

Table of Contents