Technical data Proline Promass E 500 PROFIBUS PA
286 Endress+Hauser
Export parameter settings (.csv file or PDF file, document the measuring point
configuration)
Export the Heartbeat Technology verification report (PDF file, only available with the
Heartbeat Verification →  289 application package)
Flash firmware version for device firmware upgrade, for example
Download driver for system integration
Visualize up to 1000 saved measured values (only available with the Extended
HistoROM application package→  289)
HistoROM
data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making
operation and servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored
as a backup in the device memory. This memory can be overwritten with an updated
data record, for example after commissioning.
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
HistoROM backup T-DAT S-DAT
Available data Event logbook, e.g. diagnostic events
Parameter data record backup
Device firmware package
Driver for system integration for
exporting via web server, e.g.:
GSD for PROFIBUS PA
Measured value logging ("Extended
HistoROM" order option)
Current parameter data record (used by
firmware at run time)
Indicator (minimum/maximum values)
Totalizer value
Sensor data: e.g. nominal diameter
Serial number
Calibration data
Device configuration (e.g. SW options,
fixed I/O or multi I/O)
Storage location Fixed on the user interface PC board in the
connection compartment
Can be plugged into the user interface PC
board in the connection compartment
In the sensor plug in the transmitter
neck part
Data backup
Automatic
The most important device data (sensor and transmitter) are automatically saved in the
DAT modules
If the transmitter or measuring device is replaced: once the T-DAT containing the
previous device data has been exchanged, the new measuring device is ready for
operation again immediately without any errors
If the sensor is replaced: once the sensor has been replaced, new sensor data are
transferred from the S-DAT in the measuring device and the measuring device is ready
for operation again immediately without any errors
If exchanging the electronics module (e.g. I/O electronics module): Once the electronics
module has been replaced, the software of the module is compared against the current
device firmware. The module software is upgraded or downgraded where necessary. The
electronics module is available for use immediately afterwards and no compatibility
problems occur.
Manual
Additional parameter data record (complete parameter settings) in the integrated device
memory HistoROM backup for:
Data backup function
Backup and subsequent restoration of a device configuration in the device memory
HistoROM backup
Data comparison function
Comparison of the current device configuration with the device configuration saved in
the device memory HistoROM backup