Installation Proline Promass E 500 PROFIBUS PA
24 Endress+Hauser
Orientation Recommendation
C Horizontal orientation, transmitter at
bottom
A0015590
3)
Exception:
→  8,  24
D Horizontal orientation, transmitter at
side
A0015592
1) This orientation is recommended to ensure self-draining.
2) Applications with low process temperatures may reduce the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
3) Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
If a sensor is installed horizontally with a curved measuring tube, match the position of the
sensor to the fluid properties.
1 2
A0028774
 8 Orientation of sensor with curved measuring tube
1 Avoid this orientation for fluids with entrained solids: Risk of solids accumulating
2 Avoid this orientation for outgassing fluids: Risk of gas accumulating
Inlet and outlet runs
No special precautions need to be taken for fittings that create turbulence, such as valves,
elbows or T-pieces, as long as no cavitation occurs →  25.
A0029322 A0029323
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section