Commissioning Proline Promass E 500 PROFIBUS PA
112 Endress+Hauser
Parameter Prerequisite Description Selection / User entry /
User interface
Reference sound velocity In the Select gas type parameter, the
Others option is selected.
Enter sound velocity of gas at 0 °C (32
°F).
1 to 99 999.9999 m/s
Reference sound velocity In the Select medium type parameter,
the Others option is selected.
Enter sound velocity of gas at 0 °C (32
°F).
Signed floating-point
number
Temperature coefficient sound
velocity
In the Select gas type parameter, the
Others option is selected.
Enter temperature coefficient for the
gas sound velocity.
Positive floating point
number
Temperature coefficient sound
velocity
In the Select medium type parameter,
the Others option is selected.
Enter temperature coefficient for the
gas sound velocity.
Signed floating-point
number
Pressure compensation – Select pressure compensation type. • Off
• Fixed value
• External value
• Current input 1
*
• Current input 3
*
Pressure value In the Pressure compensation
parameter, the Fixed value option is
selected.
Enter process pressure to be used for
pressure correction.
Positive floating-point
number
External pressure In the Pressure compensation
parameter, the External value option
or the Current input 1…n option is
selected.
Shows the external process pressure
value.
* Visibility depends on order options or device settings
10.6.4 Configuring communication interface
The Communication submenu guides you systematically through all the parameters that
have to be configured for selecting and setting the communication interface.
Navigation
"Setup" menu → Communication
‣
Communication
Device address
→ 112
Parameter overview with brief description
Parameter Description User entry
Device address Enter device address. 0 to 126