Operation Proline Promass E 500 PROFIBUS PA
166 Endress+Hauser
Parameter overview with brief description
Parameter Prerequisite Description User interface Factory setting
Mass flow Displays the mass flow that is
currently measured.
Dependency
The unit is taken from: Mass
flow unit parameter
(→  109)
Signed floating-point
number
Volume flow Displays the volume flow that
is currently calculated.
Dependency
The unit is taken from the
Volume flow unit parameter
(→  109).
Signed floating-point
number
Corrected volume flow Displays the corrected volume
flow that is currently
calculated.
Dependency
The unit is taken from:
Corrected volume flow unit
parameter (→  109)
Signed floating-point
number
Density Shows the density currently
measured.
Dependency
The unit is taken from the
Density unit parameter
(→  109).
Signed floating-point
number
Reference density Displays the reference density
that is currently calculated.
Dependency
The unit is taken from:
Reference density unit
parameter (→  109)
Signed floating-point
number
Temperature Shows the medium
temperature currently
measured.
Dependency
The unit is taken from:
Temperature unit parameter
(→  110)
Signed floating-point
number
Pressure value Displays either a fixed or
external pressure value.
Dependency
The unit is taken from the
Pressure unit parameter
(→  110).
Signed floating-point
number
Concentration For the following order code:
Order code for "Application
package", option ED
"Concentration"
The software options
currently enabled are
displayed in the
Software option
overview parameter.
Displays the concentration that
is currently calculated.
Dependency
The unit is taken from the
Concentration unit
parameter.
Signed floating-point
number