Proline Promass E 500 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 235
Diagnostic information Remedy instructions
No. Short text
529 Concentration settings faulty 1. Check concentration settings
2. Check input values e.g. pressure, temperature
Measured variable status
Quality Bad
Quality substatus Function check
Coding (hex) 0x3C to 0x3F
Status signal S
Diagnostic behavior Warning
Influenced measured variables
• Carrier mass flow
• Target corrected volume flow
• Carrier corrected volume flow
• Concentration
• Density
• Mass flow
• Target mass flow
• Carrier volume flow
• Target volume flow
• Volume flow
Diagnostic information Remedy instructions
No. Short text
537 Configuration 1. Check IP addresses in network
2. Change IP address
Measured variable status
Quality Good
Quality substatus Function check
Coding (hex) 0xBC to 0xBF
Status signal F
Diagnostic behavior Warning
Influenced measured variables
–
Diagnostic information Remedy instructions
No. Short text
594 Relay output simulation Deactivate simulation switch output
Measured variable status
Quality Good
Quality substatus Function check
Coding (hex) 0xBC to 0xBF
Status signal C
Diagnostic behavior Warning
Influenced measured variables
–