Proline Promass E 500 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 235
Diagnostic information Remedy instructions
No. Short text
529 Concentration settings faulty 1. Check concentration settings
2. Check input values e.g. pressure, temperature
Measured variable status
Quality Bad
Quality substatus Function check
Coding (hex) 0x3C to 0x3F
Status signal S
Diagnostic behavior Warning
Influenced measured variables
Carrier mass flow
Target corrected volume flow
Carrier corrected volume flow
Concentration
Density
Mass flow
Target mass flow
Carrier volume flow
Target volume flow
Volume flow
Diagnostic information Remedy instructions
No. Short text
537 Configuration 1. Check IP addresses in network
2. Change IP address
Measured variable status
Quality Good
Quality substatus Function check
Coding (hex) 0xBC to 0xBF
Status signal F
Diagnostic behavior Warning
Influenced measured variables
Diagnostic information Remedy instructions
No. Short text
594 Relay output simulation Deactivate simulation switch output
Measured variable status
Quality Good
Quality substatus Function check
Coding (hex) 0xBC to 0xBF
Status signal C
Diagnostic behavior Warning
Influenced measured variables