Proline Promass E 500 PROFIBUS PA Technical data
Endress+Hauser 267
Load 0 to 700 Ω
Resolution 0.38 µA
Damping Configurable: 0 to 999 s
Assignable measured
variables
Mass flow
Volume flow
Corrected volume flow
Density
Reference density
Temperature
Electronics temperature
Oscillation frequency 0
Oscillation damping 0
Signal asymmetry
Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Function Can be configured as pulse, frequency or switch output
Version Open collector
Can be set to:
Active
Passive
Passive NAMUR
Ex-i, passive
Maximum input values DC 30 V, 250 mA (passive)
Open-circuit voltage DC 28.8 V (active)
Voltage drop For 22.5 mA: ≤ DC 2 V
Pulse output
Maximum input values DC 30 V, 250 mA (passive)
Maximum output current 22.5 mA (active)
Open-circuit voltage DC 28.8 V (active)
Pulse width Configurable: 0.05 to 2 000 ms
Maximum pulse rate 10 000 Impulse/s
Pulse value Configurable
Assignable measured
variables
Mass flow
Volume flow
Corrected volume flow
The range of options increases if the measuring device has one or more
application packages.
Frequency output
Maximum input values DC 30 V, 250 mA (passive)
Maximum output current 22.5 mA (active)
Open-circuit voltage DC 28.8 V (active)
Output frequency Configurable: end value frequency 2 to 10 000 Hz(f
max
= 12 500 Hz)
Damping Configurable: 0 to 999.9 s
Pulse/pause ratio 1:1