At the time of the cycle of calibration of the feeler, the edger takes into account the wear of the feelers
since the last calibration and adjusts it to mitigate any possible variation.
Have the plastic calibration gauge to hand.
Press on > > > to reach the menu of calibration.
The following screen is displayed:
Place the calibration gauge in the posiblock holder.
Press to tighten the clamps.
Press to start the calibration cycle.
The gauge is positioned at the level of the feeler, the feeler moves.
Does a message appears “is the point of the feeler positioned correctly in the drill-hole of the plastic
pattern?”
Check that the point of the feeler is positioned correctly.
Press if the feeler position is correct.
>
>
1
2
3
4
5
Delta - Edging System > V1- 02-2020
154
USER MANUAL> IX. MAINTENANCE AND SERVICING