20 UMG70XV3-4EN 12.01.2009
2 Safety Instructions according to Directive 94/9/EC
Type **K1N43, **G1N43, **H1N43, **M1N43, **P1N43
manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin
marking
0044;
II2G Ex q II T6…T3 T
a
-55…+140 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
type of protection transducer: powder filling
dust: protection by enclosure tD
degree of protection IP 65 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T3: -55…+140 °C
temperature class T4: -55…+125 °C
temperature class T5: -55…+90 °C
temperature class T6: -55…+75 °C
Type **M2E45, **P2E45
manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin
marking
0044;
II2G Ex q II T6…T2 T
a
-45…+225 °C
II2D Ex tD A21 IP65 TX
certification IBExU04ATEX1011 X
type of protection transducer: powder filling
dust: protection by enclosure tD
degree of protection IP 65 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T2: -45…+225 °C
temperature class T3: -45…+190 °C
temperature class T4: -45…+125 °C
temperature class T5: -45…+90 °C
temperature class T6: -45…+75 °C
The service temperature at the cable entry must not
exceed the permissible operating temperature of 190 °C.