UMG70XV3-4EN 12.01.2009 21
2 Safety Instructions according to Directive 94/9/EC
Type **K1N71, **G1N71
manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin
marking
0044
Ex e q II T6...T3
Ex tD A21 IP65 TX
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
type of protection transducer: powder filling
transducer connection: increased safety
dust: protection by enclosure tD
degree of protection IP 65 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T3: -40…+180 °C
temperature class T4: -40…+125 °C
temperature class T5: -40…+90 °C
temperature class T6: -40…+75 °C
Type **K1N81, **G1N81, **M2N81, **P2N81, **Q2N81
manufacturer FLEXIM GmbH, Wolfener Str. 36, 12681 Berlin
marking
0044;
Ex e q II T6...T3
Ex tD A21 IP65 TX
certification IBExU07ATEX1168 X
IECEx IBE08.0007X
type of protection transducer: powder filling
transducer connection: increased safety
dust: protection by enclosure tD
degree of protection IP 65 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T3: -55…+180 °C
temperature class T4: -55…+125 °C
temperature class T5: -55…+90 °C
temperature class T6: -55…+75 °C
II2G
II2D
II2G
II2D