UMG70XV3-4EN 12.01.2009 23
2 Safety Instructions according to Directive 94/9/EC
Type **K1LI1, **G1LI1, **M2LI1, **P2LI1
manufacturer
FLEXIM GmbH
Wolfener Str. 36
12681 Berlin
marking
0044; II 2G 2D
Ex q II T6...T3
Ex tD A21 IP68 TX
certification
IBExU07ATEX1168 X
IECEx IBE08.0007X
type of protection transducer: powder filling
dust: protection by enclosure tD
degree of protection
IP 68 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T3: -55…+180 °C
temperature class T4: -55…+125 °C
temperature class T5: -55…+90 °C
temperature class T6: -55…+75 °C
The ultrasonic transducers may be used at the cable
gland only in the temperature range of -40...+80 °C.
Type **M2E85, **P2E85, **Q2E85
manufacturer
FLEXIM GmbH
Wolfener Str. 36
12681 Berlin
marking
0044;
Ex e q II T6...T2
Ex tD A22 IP56 TX
certification
IBExU07ATEX1168 X
IECEx IBE08.0007X
type of protection transducer: powder filling
transducer connection: increased safety
dust: protection by enclosure tD
degree of protection
IP 56 according to EN 60529
explosion protection
temperature according
to ATEX
temperature class T2: -45…+225 °C
temperature class T3: -45…+190 °C
temperature class T4: -45…+125 °C
temperature class T5: -45…+90 °C
temperature class T6: -45…+75 °C
II2G
II3D

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS G704 and is the answer not in the manual?

Flexim FLUXUS G704 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS G704
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals