FT-111, User Manual Rev. 2.0.0, June 2022 Page 83 of 137
Set/Reset digital outputs
Read weight value in
increased resolution
[ADR][X][STATUS][SIGN][WEIGHT
VALUE]
BSI-Base Commands and Responses:
Read all weight data immediately
Read Gross weight value immediately
Read the last printed data
Read stable weight and Record to Alibi
Erase the logs in the SD card.
Start /stop continuous data output
Read voltage value of DC power supply
Read current weight (indicated) value immediately
Read the logs in the SD card.
Read the current stable weight value
Set/Reset digital outputs
Read current weight value in increased resolution immediately
Command : [ADR][A]
Response : [ADR][A][STATUS][SIGN][NET W][SIGN][TARE W][SIGN][GROSS W]
Example :
Command : 01A
Response : 01AS+000123.4+000111.1+000234.5
01AD+000123.4+000111.1+000234.5
01AO (ADC out error )
Comments :
The response is net, tare and gross weight values or error status.
All weight data is transmitted immediately after receiving the command.
Command : [ADR][B]
Response : [ADR][B][STATUS][SIGN][WEIGHT VALUE]
Example :
Command : 01B
Response : 01BS+000123.4 (gross weight is stable and 123.4)
01BD+000123.4 (gross weight is dynamic and 123.4)
(under load)
Comments :
The response is the gross weight value (stable or dynamic) or error status.
Gross weight data is transmitted immediately after receiving command.