SECTION 3: Troubleshooting
2025653-048 Revision B Responder
2000 Page 25
Shock into Patient Fails
1. Perform a button test for each Paddle.
If the test fails, replace the Paddle.
If the test does not fail, connect the Paddles/Pads/Spoons to a simulator and attempt to shock.
2. If the shock is successful:
The unit is working correctly. Check patient preparation.
3. If the shock is unsuccessful:
1. Replace the Paddles/Pads/Spoons.
2. Check the cables between the Rear Panel and the Therapy Board.
3. Check the cable between the Therapy Board and the Main Board.
4. Replace the Therapy Board.
5. Replace the Main Board.
No ECG from Paddles/Pads/Spoons
1. Replace the paddles/pads/spoons.
2. Check the cables between the Rear Panel and the Therapy Board.
3. Check the cable between the Therapy Board and the Main Board.
4. Replace the Therapy Board.
5. Replace the Main Board.
Service Required LED is ON
If the Service Required LED is on and the fault has been corrected, perform this procedure to clear the fault.
1. Go to Maintenance > Stored Data Management > Clear Service Required Indicator.
2. Press Accept.
3. Cycle power.
If the fault does not clear, go to History > Event Log to find the fault log entry.
Note: All service required Error codes are displayed in the History with a red exclamation mark.

Table of Contents

Other manuals for GE Responder 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Responder 2000 and is the answer not in the manual?

GE Responder 2000 Specifications

General IconGeneral
BrandGE
ModelResponder 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals