Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 67
ECG Lead Test
1. Connect the ECG leads to the patient simulator.
2. On the Responder 2000, select ECG Lead Test
from the Maintenance menu.
3. Verify ECG Lead Status display as shown:
4. Disconnect RA (White Lead) from Patient Simulator
and verify the display as shown:
5. Reconnect RA lead to patient simulator.
6. Disconnect RL (Green Lead) from Patient Simulator
and verify the display as shown:
7. Reconnect RL lead to patient simulator.
8. Disconnect LA (Black Lead) from Patient Simulator
and verify the display as shown:
9. Reconnect LA lead to patient simulator.
10. Disconnect LL (Red Lead) from Patient Simulator
and verify the display as shown:
11. Reconnect LL lead to patient simulator.