EasyManua.ls Logo

GE Responder 2000

GE Responder 2000
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION 8: Specifications and Safety
2025653-048 Revision B Responder
2000 Page 96
System Overview
The Responder 2000 system has the following functions and features.
Note: Refer to the Operator’s Manual for specific instructions and intended use.
Functions
Defibrillation (with dedicated Defibrillator operation buttons)
2J to 270J Biphasic waveform (can deliver 270J Biphasic shock into a 50 Ohm load)
Paddle connector accepts pads, paddles, or spoons
Transcutaneous cardiac pacing
40ms rectilinear current source pace pulse, 0 to 140 mA, 0 to 180 PPM, demand and asynchronous modes
ECG Monitoring through therapy electrodes, paddles, or a separate five wire ECG input
Optional Pulse Oxymetry (GE506)
Features
Color LCD to display operational, ECG, and SPO
2
information
The display is 5.67”, backlit, with a 320 x 240 resolution and over 260k colors
Trim knob for input via the display
Thermal printer for ECG, Pulse Oxymetry, Therapy Actions and Event history
384 pixel resolution at 25mm/s on a 60mm paper roll
Prints in real-time two-trace graphic, text, and raster modes
AC or rechargeable battery operation
AC: 90 VAC to 264 VAC, 47Hz-63Hz, 90 Watts Cont., 115 Watts intermittent
Battery: Lithium Ion with built in charge indicator, delivers up to 10A at 11.0V (Nominal)
Internal Defibrillator and Pacer test load
Speaker for audible indication
Serial port for software upgrade

Table of Contents

Other manuals for GE Responder 2000

Related product manuals