Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 57
Temperature Readings
Displays internal temperatures for the Power Board, Therapy Board, Printer, and Battery. These readings may vary
according to fan speed or external conditions. An abnormally high reading may indicate a fan not working or other fault.
The internal fan should turn on at the following temperatures:
Fan ON (High) Fan ON (Low) Fan OFF
Power Board
65º 60º 55º
Therapy Board
65º 60º 55º
Printer
N/A N/A N/A
Battery
N/A N/A N/A
Fan Test
Select Off, Low, or High
Note: The Low fan speed may be difficult to hear in a noisy environment.