SECTION 4: Repair
2025653-048 Revision B Responder
™
2000 Page 30
Display/Main Board/Printer (see page 42)
1. Follow the Capacitor, Therapy Board, SPO2 Board, and ECG Board instructions (above) to remove the
Capacitors Therapy Board, SPO2 Board, and ECG Board.
2. Remove the screws holding the Front Bezel (see Figure 45: Front Bezel Installed).
3. Remove the screws holding the Display (see Figure 43: LCD Screw Locations).
At this point the Display can be replaced (disconnect Display cables).
If the Main Board or Printer must be replaced, perform the following steps:
Main Board
a. Remove the screws (see Figure 11: Main Board Installed).
b. Disconnect all cables
c. Exchange the Main Board.
d. Reconnect all cables.
e. Replace the screws.
Printer
a. Remove the screws (see Figure 14: Printer Screws).
b. Disconnect the ribbon and Power cables.
c. Exchange the Printer.
d. Reconnect ribbon and Power cables.
Note: Removing the printer bracket may help when connecting the Printer ribbon cable to the Main Board
(see Figure 10: Printer Bracket Removal).
e. Replace the screws.
4. Connect the Display cables.
5. Replace the Display screws.
6. Follow the ECG Board, SPO2 Board, Therapy Board, and Capacitor instructions (above) to replace the ECG
Board, SPO2 Board, Therapy Board, and Capacitor.
Power Supply and Power Control Board (see page 36)
Remove the Base Cover, disconnect all cables, and remove screws.
Note: The Power Supply is held in place only by the two larger Base Cover screws.
Switches and Fan (see page 36)
Remove Base Cover and Power Control Board.