Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 60
Visual Inspection
Visually inspect the unit and accessories for foreign substances, damage, cracks, bent or discolored pins, broken parts, or
extreme wear, including:
Cracked, bulging, or otherwise damaged case
Cracked, loose, or bent connectors or plugs
Rattling when gently shaken
Discoloration or signs of overheating
Obvious scratches or damage to the display
Loose battery or paddle clips
Loose printer door
Operational Tests
Initial Power Up
Perform this procedure for the following scenarios:
With a charged battery (at least 25%) installed and connected to AC power
With the battery removed and connected to AC power
With a charged battery (at least 25%) and AC power disconnected
Turn on the Responder 2000 and verify the following during power up:
No alarms (at any time during power up or operation)
The speaker emits a short beep, then is silent
The following indicators momentarily light:
o Service Required: Red
o Charge: off
o Manual: Blue
o Shock: Red
The Responder 2000 is ready for use in about 4 seconds.
The AC Power LED and Battery Charging LED are lit when a battery is installed and the Responder 2000 is connected to
AC power
The AC Power LED is not lit and Battery Charging LED is not lit when a battery is installed and the Responder
2000 is not connected to AC power

Table of Contents

Other manuals for GE Responder 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Responder 2000 and is the answer not in the manual?

GE Responder 2000 Specifications

General IconGeneral
BrandGE
ModelResponder 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals