Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
™
2000 Page 60
Visual Inspection
Visually inspect the unit and accessories for foreign substances, damage, cracks, bent or discolored pins, broken parts, or
extreme wear, including:
• Cracked, bulging, or otherwise damaged case
• Cracked, loose, or bent connectors or plugs
• Rattling when gently shaken
• Discoloration or signs of overheating
• Obvious scratches or damage to the display
• Loose battery or paddle clips
• Loose printer door
Operational Tests
Initial Power Up
Perform this procedure for the following scenarios:
• With a charged battery (at least 25%) installed and connected to AC power
• With the battery removed and connected to AC power
• With a charged battery (at least 25%) and AC power disconnected
Turn on the Responder 2000 and verify the following during power up:
• No alarms (at any time during power up or operation)
• The speaker emits a short beep, then is silent
• The following indicators momentarily light:
o Service Required: Red
o Charge: off
o Manual: Blue
o Shock: Red
• The Responder 2000 is ready for use in about 4 seconds.
The AC Power LED and Battery Charging LED are lit when a battery is installed and the Responder 2000 is connected to
AC power
• The AC Power LED is not lit and Battery Charging LED is not lit when a battery is installed and the Responder
2000 is not connected to AC power