Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 69
Safety Tests
WARNING: Explosion or Shock Hazard.
Before performing any test, read and follow all safety instructions for the test equipment indicated.
WARNING: Explosion or Shock Hazard.
The following instructions are written for the test equipment indicated. Performing any test with equipment other than the
specific brand and model indicated may cause explosion, electrical shock, and equipment damage.
All safety tests are required whenever the case is opened, regardless of the type of maintenance or service performed.
Safety tests include:
Hipot Test
Leakage Current Test
Note: If the Hipot and Leakage current tests in this section cannot be performed for any reason, perform the Alternate
Safety Test Procedures starting on page 75.
Hipot Test
WARNING: Explosion Hazard.
Internal damage or improper assembly could cause an internal component to explode during Hipot testing. To minimize
risk
Ensure all Responder 2000 covers are in place and securely attached before testing
Use appropriate personal safety equipment, including a face shield and gloves
Ensure the area is clear before starting the test.
WARNING: Shock Hazard.
Do not touch the Hipot tester, the Responder 2000, or any leads during the test. Ensure all Hipot leads are electrically
isolated before starting the test. If the test is performed on an anti-static mat, ensure the leads do not contact the mat
during the test.
Required Equipment
UadTech Sentry 30 AC/DC/IR Hipot Tester
Test fixtures
o ECG Connector Test Fixture. ECG connector with all leads and ground shield shorted together.
o AC Plug Test Fixture. AC connector with line and neutral shorted together.
o SpO
2
Connector Test Fixture. Connection for the ground shield.
o ECG Paddle Test Fixture. Apex and Sternum pins shorted together.
o Serial Port Test Fixture. Serial port connector with single lead connect to pin 5.

Table of Contents

Other manuals for GE Responder 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Responder 2000 and is the answer not in the manual?

GE Responder 2000 Specifications

General IconGeneral
BrandGE
ModelResponder 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals