Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
™
2000 Page 51
Required Equipment
• Fluke Model Index 2 Sp02 simulator or (equivalent)
• Fluke Model Impulse 4000 Defib analyzer (or equivalent)
• Metron QA-90 Safety Analyzer (or equivalent)
• UadTech Sentry 30 AC/DC/IR Hipot Tester
PRECAUTION: Do not shock into devices with highly non-linear resistance versus the voltage, (e.g. an overvoltage
protector). This may damage the Responder 2000. Do not use the HELLIGE Test Lamp 301 495 00 for testing.
This test lamp includes an overvoltage protector. Shocking into it may damage the Responder 2000.
HELLIGE Test Lamp
Preventive Maintenance
The following preventative maintenance and operational checks are normally performed only by the user:
• Cleaning the Responder 2000 or accessories
• Calibrating the battery fuel gauge
• Daily maintenance per Defibrillator Checklist
• Other routine maintenance
Refer to the Operator’s Manual for more information.
Annual Inspection
A qualified service technician should check the Responder 2000 once per year for the following:
1. Visual Inspection (page 59)
2. Operational Check (starting on page 60)
3. Printer Test (page 68)