Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 70
ECG to Serial 4.0 KVAC Test
1. Setup the Hipot tester:
4.0 KVAC
High current limit 4.0 mA
Test 30 Sec.: Ramp 2.0 Sec.
120 VAC (±10 %) Input
2. Plug in the ECG Connector Test Fixture into the ECG port.
3. Plug the Serial Port Test Fixture into the Serial port.
4. Connect the red hipot tester lead to the ECG Connector Test Fixture lead.
5. Connect the black hipot tested lead to the Serial Port Test Fixture lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
6. Press Start (green button) on the Hipot tester. The test will take approximately 30 seconds.
If the test passes (short beep and green Pass LED is lit), continue with the next test.
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.
AC Line/Neutral to AC Ground 1.5 KVAC Test
1. Setup the Hipot tester:
1.5 KVAC
High current limit 4.0 mA
Test 30 Sec.: Ramp 2.0 Sec.
120 VAC (±10 %) Input
2. Plug in the AC Plug Test Fixture to the AC input connector.
3. Connect the red hipot tester lead to the AC Plug Test Fixture Line and Neutral lead.
4. Connect the black hipot tested lead to the AC Plug Test Fixture Ground lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
5. Press Start (green button) on the Hipot tester. The test will take approximately two minutes.
If the test passes (short beep and green Pass LED is lit), continue with the next test.
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.
ECG to SpO
2
1.5 KVAC Test
Note: This test is only performed on models with the SpO
2
option installed.
1. Setup the Hipot tester:
1.5 KVAC
High current limit 4.0 mA
Test 30 Sec.: Ramp 2.0 Sec.
120 VAC (±10 %) Input
2. Plug in the ECG Connector Test Fixture into the ECG port.
3. Plug the SpO
2
Test Fixture into the Serial port.
4. Connect the red hipot tester lead to the SpO
2
Test Fixture lead.
5. Connect the black hipot tested lead to Serial Port Test Fixture lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
6. Press Start (green button) on the Hipot tester. The test will take approximately two minutes.
If the test passes (short beep and green Pass LED is lit), continue with the next test.
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.

Table of Contents

Other manuals for GE Responder 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Responder 2000 and is the answer not in the manual?

GE Responder 2000 Specifications

General IconGeneral
BrandGE
ModelResponder 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals