What to do if Operating System Error appears on GE Sievers Innovox Measuring Instruments?
- ZZachary MaldonadoSep 23, 2025Please power cycle the Analyzer if your GE Measuring Instruments displays an Operating System Error. 

What to do if Operating System Error appears on GE Sievers Innovox Measuring Instruments?
Please power cycle the Analyzer if your GE Measuring Instruments displays an Operating System Error.
What should I do if my GE Analytical Instruments displays Stream 1 Flow Warning?
Check that water is flowing in the sample cup and the inlet valve (if installed) is properly adjusted. If the problem persists, please export the warnings and contact GE Analytical Instruments for further instructions.
What to do if my GE Analytical Instruments displays a Fluid Leak in Optional C1D2 or ATEX Enclosure Error?
This error indicates that a large leak has been detected inside the optional outer enclosure. You should resolve the source of the leak and then clean and dry the sensor.
How to troubleshoot Reactor Gas Flow Error on GE Sievers Innovox?
To address a Reactor Gas Flow Error on your GE Measuring Instruments, go to the Diagnostics screen and run the Gas Valves test (V2-R). If the gas flow rate fails to maintain 165 cc/min, there may be an issue.
What to do if iButton Read Error appears on GE Sievers Innovox?
Please power cycle the Analyzer if your GE Measuring Instruments displays an iButton Read Error.
How to resolve Operating System Error on GE Analytical Instruments?
This is a 'catch all' for any errors not specifically defined elsewhere. Please power cycle the Analyzer. If the problem persists, export errors and contact GE Analytical Instruments for further instructions and analysis.
What to do if my GE Analytical Instruments displays a Datastore Error?
An error has been found during writing to the data history portion of the flash memory card on the Analyzer’s System Controller Board. Please shut down the Analyzer and power cycle. If the problem persists, please export the errors and contact GE Analytical Instruments for further instructions.
How to troubleshoot Carrier Bypass Gas Flow Error on GE Measuring Instruments?
To address a Carrier Bypass Gas Flow Error on your GE Measuring Instruments, go to the Diagnostics screen and run the Gas Valves test (V1-R). If the gas flow rate fails to maintain 165 cc/min, there may be an issue.
Why does my GE Analytical Instruments show a Peristaltic Pump Life Error?
This error appears because, based on its usage history, the analyzer estimates that the peristaltic pump in the Sample Handling and Reactor modules has reached the end of its life. To resolve this, order two new peristaltic pumps from GE.
What to do if my GE Analytical Instruments displays a Touchscreen Error?
Please shut down the Analyzer and power cycle. If the problem persists, shut down the Analyzer and power down. Then, check the cable connections. You may be able to regain control of the unit by plugging a mouse into any of the USB port. If the problem persists please export the errors and contact GE Analytical Instruments for further instructions.
| Category | Measuring Instruments | 
|---|---|
| Brand | GE | 
| Model | Sievers Innovox | 
| Response Time | Typically 2 minutes | 
| Power Supply | 100-240 VAC, 50/60 Hz | 
| Communication Interfaces | USB, Ethernet | 
| Certifications | CE | 
| Humidity Range | 10% to 90% non-condensing | 
Provides detailed technical specifications for the InnovOx TOC Analyzer.
Discusses sample species that can form toxic products or react violently.
Details interferences that can affect component longevity and performance.
Explains how inorganic compounds can affect measurement accuracy.
Procedure for connecting the waste tube to a drain.
Details on connecting various cables and I/O interfaces.
Instructions for safely connecting the analyzer to AC power.
Guidance on connecting sample streams to the analyzer.
Procedures for installing the wastewater sampler.
Guides initial power-up and basic analyzer configuration.
Procedure to confirm the analyzer is operating correctly.
Step-by-step guide to performing the calibration process.
Instructions for reviewing and modifying calibration measurements.
Defines measurement settings for Online, Check Standard, Grab, and Rinse modes.
Guides on setting up online measurements for various features.
Details parameters for NPOC, TOC, TC, and IC measurements.
Procedure for measuring carbon concentrations in reagents for blank correction.
Guides on configuring analyzer interfaces and connections.
Steps to manually configure IP settings for network connection.
Safe procedures for powering down the analyzer.
Procedure for lubricating the analyzer's syringes every three months.
Six tests to locate leaks in the gas delivery system.
Correction for carbon contribution from reagents.