What does it mean when my GE Analytical Instruments displays a Gas Flow Regulation Error?
- Mmichelle29Aug 29, 2025
This error means that the PID Test has either failed to stabilize in the allotted time or is outside the acceptable range. To resolve this: 1. Determine which Gas Valve PID settings should be selected by identifying which Gas Manifold is installed in your instrument. 2. Verify the appropriate PID option is activated in the Analyzer (Maintenance > Advanced). 3. Perform the Gas Leak Check diagnostics to identify any other possible problems. If the error continues, contact GEAI Technical Support for guidance.