GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-99
3-10-3-3 Set the Scanner’s Network Information
In the IP settings area of the screen, enter the following:
1.) IP Address for scanner. (Default IP Address from factory: 10.0.0.3).
2.) Subnet Mask for scanner. (Default Subnet Masks from factory: 255.255.255).
3.) IP address for Default Gateway. (Default Gateways from factory IP address: 0.0.0.0).
4.) In addition, the scanner’s AE Title must be entered in the DICOM server’s setup.
.
3-10-3-4 Set the Remote Archive’s Network Information
In the Remote Archive Setup area of the screen (see example in Figure 3-102), enter the following:
1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Archive Name from factory: ECHOPAC7-000001).
.
Figure 3-101 TCP/IP Set-up for Vivid i n/ Vivid q N
Figure 3-102 TCP/IP Set-up for Vivid i n/ Vivid q N
IP Address
Subnet Mask
Default Gateway
AETitle
IP Address for the remote archive
Computer Name for the remote
archive