GE HEALTHCARE
DIRECTION FQ091013, REVISION 1    VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting  7-97
• Yellow: The tests executed so far on these components have passed, but not all 
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Temperatures Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details 
of which tests passed or failed, as shown in the example in Figure 7-67.
Note: For this test, there is one report (Temperatures) which is listed in the Reports window on 
the right. By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values are within the pre-defined Max and Min limits, as 
shown in Figure 7-67.
8) When done, trackball to the Exit button and press Set to close the View Report window.
Figure 7-67   View Test Log - Temperatures Test