DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-167
3.) If use of the Advanced features is required, proceed to the following instructions, as applicable:
Enabling Extensive Logging on page 7 - 169
Advanced Options on page 7 - 171
DB Scan on page 7 - 174
After the required advanced features have been selected, continue with step 4 and step 5, below.
4.) From the Destination drop-down menu, select the required export destination (DVD/CD or Memory
Stick) then click Save and Export.
A Zip file (named "logfile_<date>_<time>.zip") is copied onto the selected removable media.
Note: Previously, users were required to save log files before exporting them. With version BT06,
these two operations are performed as one step.
5.) When done, click the Exit button.
The System Problem Reporting dialog closes.
6.) Press <Ctrl>+E to eject the media.
Figure 7-144 System Problem Reporting Dialog
When selecting DVD/CD as the export destination, first make sure that the DVD/CD has been
formatted. For details, see Formatting a DVD/CD on page 7 - 168.