GE HEALTHCARE
DIRECTION FQ091013, REVISION 1    VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting  7-171
7-8-2-3 Advanced Options
1.) From the System Problem Reporting dialog (Figure 7-144 on page 7-167), click the Options button.
The Advanced Export Options screen opens as shown in Figure 7-148.
Note: The Advanced Export Options allows you to:
a.) Control the size of log files.
b.) Specify some optional attachments to be added to the log files.
2.) Proceed to select the appropriate settings, as described below.
7-8-2-3-1 Setting the Log File Ranges
Note: By default, the log files may be very large as they are not limited by time/date. The user is 
able to limit the size/timeframe of the log files by choosing the required settings, as 
described below.
1.) Activate the Export Logs Using checkbox, as shown in Figure 7-148, above.
2.) Proceed to set the required Log Range, as follows:
Either:
Activate the Time Range radio button, then use the From:/To: drop-down lists to define the required 
time range.
Or:
Activate the Bookmark radio button, then select the required Bookmark from the drop-down list of 
bookmarks.
This will export log files which are limited to the range of 1 hour before through 1 hour after the 
selected bookmark.
Figure 7-148   Advanced Export Options Dialog