GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
7-170 Section 7-8 - Automatic Error Log
1.) Perform step 1 and step 2 on page 7-166.
2.) Click the Extensive Log button (refer to Figure 7-144 on page 7-167).
The Extensive Logging dialog opens - see Figure 7-147, below.
3.) Flag the appropriate checkboxes to activate the categories for which detailed analysis is required
(refer to the recommendations in Table 7-4).
Note: The impact of activating a checkbox would be to generate larger, detailed log files.
After usage, make sure to return to the default settings (unflagged) to conserve the size of
the log files. After 7 days, the system will automatically return to the default settings.
4.) When done, click OK.
The Extensive Logging dialog closes.
Figure 7-147 Extensive Logging Dialog