DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-30 Section 8-3 - Internal Component Replacement Procedures
8-3-1-4 P&M (Probe and MUX) Board Removal Procedure
1) Remove the bearing handle, as described in the Bearing Handle Removal Procedure on page 8 - 3.
2) Remove the Control Panel and Keyboard, as described in the Control Panel and Keyboard
Removal Procedure on page 8 - 9.
3) Loosen and remove the four screws that secure the P&M Board to the RFI Board, as shown in
Figure 8-33.
4.) Loosen the two screws that fasten the P&M Board to the Bottom Assembly, as shown in
Figure 8-34 below.
Figure 8-33 Removing the Screws that Secure the P&M Board to the RFI Board
Figure 8-34 Loosening the Screws from the P&M Board