GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-60 Section 8-3 - Internal Component Replacement Procedures
5) Using both hands, carefully lift the LCD Display upwards and install it back to the Rear Cover &
Latch Assembly, see Figure 8-68 on page 8-58.
6.) Connect the flex cable to the BEP.
7) Return the LCD Display as described in the LCD Display Installation Procedure on page 8 - 55.
8) Return the LCD Display Frame, as described in the LCD Display Frame Installation Procedure on
page 8 - 28.
9) Check that the LCD Display can be tilted upwards and downwards, and that it holds its set position.
10) Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation
Procedure on page 8 - 11.
11.)Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23 - only if the Vivid i n/ Vivid q N system
is currently mounted on the SafeLock cart
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
- Full System Test on page 7 - 10
Figure 8-71 Adhering Kapton Tape on the LCD Rear Panel Connector