GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-77
9.) Replace all tie-wraps.
10.)Shorten the USB power cable by folding it into a "u" shape and fasten together with tie wraps as
shown in Figure 8-91.
11.)Check that the routing procedure has been implemented correctly: lower the cart’s top-shelf and
make sure that the cables do not interfere with the movement of the cart’s top shelf.
Figure 8-90 Peripheral AC Outlet and Ground Cable Connected
Figure 8-91 USB Power Cable Folded and Fastened with Tie-wraps