8.) Scroll to Memory Cache and press <Enter> to open the Memory Cache screen.
9.) Set the parameters in the Memory Cache screen as shown above in Figure 8-112.
10.)Press <Esc> to return to the Advanced menu home screen (Figure 8-109).
11.)Scroll to I/O Device Configuration and press <Enter> to open the I/O Device Configuration
screen.
12.)Set the parameters in the I/O Device Configuration screen as shown above in Figure 8-113.
13.)Press <Esc> to return to the Advanced menu home screen (Figure 8-109).
Figure 8-112 BIOS Advanced Memory Cache Parameters
Figure 8-113 BIOS Advanced I/O Device Configuration Parameters