DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-136 Section 8-5 - Software Loading
9) Run PMCbootloader.exe by double-clicking on it - see Figure 8-132 below.
Figure 8-132 Run PMCbootloader.exe
A window is displayed showing the current PMC version and the path of the PMCJPS2.hex file:
10.)Click Burn.
The burning process will take approximately 30 seconds.
When done, a brief message will appear “Burning process failed…” - see Figure 8-133 below.
Note: The failed message is normal and should be anticipated.
Figure 8-133 Burning Process Failed Message