DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-137
11) Click OK.
The system should shutdown automatically.
(The screen will appear blank but fans will remain running and the On/Off key will be disabled).
12) Power-down the system by removing the AC adapter.
13.)Remove the Service Dongle.
14.)Wait approximately 5 seconds, then boot-up the system again.
Allow the system to boot up into Normal Scanning mode.
15) Press Config.
The system will prompt you to login.
16) Login as ADM.
The default password is ulsadm
17) Press the About button and activate the FwVersion tab; verify that PMC version is now 23, as
shown in the example in Figure 8-134 below.
Figure 8-134 Checking the PMC Version
18) In the event that the PMC version is other than 23, burning has failed and it is necessary to repeat
the procedure from step 5 above. Also, refer to the note below.