83
6. When sample is not spotted onto slides.
Error Code Error Message Error Condition Troubleshooting
Sample application problem <, @ or ¥ . Because the sample is not
applied to the slide, the
error mark <, @ or ¥ appears
continuously.
IMPORTANT
Measurement values with an
[@] mark might be incorrect.
Run the measurement again.
Check the following:
a. Make sure that DRI-CHEM Auto Tips are
being used.
b. Make sure that tips are not being reused.
c. Make sure that there are no bubbles on the
sample surface in the sample tube.
d. Check the sample for fibrin.
e. Make sure that a sample rack that matches
the sample tube size is being used.
f. Inspect the sampler O-ring. Section 6.6.1
3. When there are insufficient number of or no mixing cups.
Error Code Error Message Error Condition Troubleshooting
W0151 Lack of mixing cup.
Set mixing cup and re-run the tests from the
beginning.
There are not enough mixing
cups.
After you replace the mixing cups, re-run
from the measurement request. Section 5.3
W0152 No mixing cup is detected.
Set mixing cup and re-run the tests from the
beginning.
No mixing cups can be detected.
After you replace the mixing cups, re-run
from the measurement request. Section 5.3
W0154 Mixing cup detection sensor ERR Error of the mixing cup detection
sensor.
If the error still occurs after switching the
analyzer off and on, please contact Heska
Technical Support Services.
4. During a sample volume error.
Error Code Error Message Error Condition Troubleshooting
W0120 Sample shortage
Sample is insufficient. Check the sample.
Either the sample is not loaded or
the sample volume is insufficient.
Make sure that the sample volume in the
tube is within the measurable range and
an appropriate sample tube is used in
the sample rack, and then rerun from the
measurement request.
NOTE: If the sample volume is low, add
sample to the tube or perform manual
sample application.
W0060 Sample shortage
Check the sample volume.
Confirm that a sample rack matching the
sample tube size is used.
There is not enough sample
volume.
W0140 Surface detection ERR (sample)
Sample volume is too much. Check the sample
volume.
There is too much sample
volume.
5. During a sample aspiration or application error.
Error Code Error Message Error Condition Troubleshooting
E0111 Aspiration clogging ERR (sample)
Clogging was detected during sampling. Check
the sample.
Clogging was detected during
sample aspiration.
The sample may contain fibrin deposits.
After checking the sample, rerun from the
measurement request.
E0124 Aspiration clogging ERR (diluted sample)
Clogging was detected during sampling. Check
the sample and diluent.
Clogging was detected during
sample aspiration.
E0112 Spotting clogging ERR (sample)
Clogging was detected during sampling. Check
the sample.
A clog was detected during
sample application (ejection).
E0125 Spotting clogging ERR (diluted sample)
Clogging was detected during sampling. Check
the sample and diluent.
A clog was detected during
sample application (ejection).
The sample may contain fibrin deposits.
After checking the sample, rerun from the
measurement request.
NOTE: If the error occurs again, there might
be a piping system problem. Please contact
Heska Technical Support Services.