Fig. 140: Main screen active
1 Rotating circle symbol
A dened pr
ogram runs in program mode.
On the main screen, the rotating circle symbol, together with the
symbol for manual mode, are displayed to indicate that temperature
control is active. The number in the circle is the currently running pro-
gram segment number of the program.
While program mode is active, you can create, edit, or delete a pro-
gram in parallel.
NOTICE
No other program can be run in parallel with program
mode.
The settings for program mode are made in the
‘Program mode set-
tings’
menu
Chapter 6.12.1 ‘Program mode settings’ on page 69
.
NOTICE
A potential touchscreen failure is indicated by the
standstill of the rotating circle symbol. The output of
an acoustic alarm is possible by connecting a warning
device to the potential-free alarm contact.
If the incubator is switched off while manual or pro-
gram mode is running, when the incubator is
switched on, it indicates that a power failure was
registered.