Conguring Alarms
> >
2
2
3
3
1
1
Congure each alarm channel ([ALM1] to [ALM8]).
1
Select the waveform display color.
× (OFF), 24 colors
2
Under [Filter] , select the number of data points.
The system will output an alarm if the alarm state continues for the set number of data points.
OFF
, 2, 5, 10, 20, 50, 100, 200, 500, 1000, User
If alarm sources include a wireless module’s channel, no lter may be applied during data recovering
processes.
(When [Filter] is set to [
User
])
Set the number of data points.
2
to 1000
3
Under [Comment], enter a comment (as necessary).
See “Text entry method” (p. 10).