362
Measurement range, maximum resolution, measurable range, and measurement accuracy gures
Measurement
target
Type Range
Maximum
resolution
Measurable range
Measurement
accuracy
Voltage – 100 mV f.s. 5 µV ±100 µV
200 mV f.s. 10 µV ±200 µV
1 V f.s. 50 µV ±1 mV
2 V f.s. 100 µV ±2 mV
10 V f.s. 500 µV ±10 mV
20 V f.s. 1 mV ±20 mV
100 V f.s. 5 mV ±100 mV
1-5
V f.s. 500 µV 1 V to 5 V ±10 mV
Temperature
characteristics
Normal-mode
rejection ratio
50 dB or greater
interval of 5 s)
interval of 5 s)
Common-mode
rejection ratio
Signal source resistance of 100
or less
100 dB or greater
(With 50 Hz/60 Hz input and a data refresh interval of 1 ms)
140 dB or greater
interval of 5 s in the 100 mV f.s.)
interval of 5 s in the 100 mV f.s.)
Eect of radiated
radio-frequency
electromagnetic eld
range)
Eect of conducted
radio-frequency
electromagnetic eld
3. Functionality specications
LED indication Wireless connection, measurement status, error status, AC Adapter- / external power
supply-powered operation, battery-powered operation, battery-charging status
Operation keys Auto, reset
Auto-connect
function
Yes