Brevera Breast Biopsy System User Guide
Chapter 2: General Information
Page 12 MAN-04303-002 Revision 006
Warning:
Cables and cords can pose a tripping hazard. Position cords safely out of the
way. When not in use, wind cords securely around the cord management
plates.
Caution:
If the console will not be used for several days or an extended period of time,
disconnect the console from the electrical outlet. Place the console in a location where
it will not be damaged.
Caution:
The console can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, can cause harmful interference to other devices in
the vicinity. There is no guarantee that interference will not occur in a particular
installation. If the console does cause harmful interference to other devices, which
can be determined by turning the console off and on, the user is encouraged to try to
correct the interference by one or more of the following measures: reorient or relocate
the receiving device; increase the separation between the equipment; connect the
console to an outlet on a circuit different from that to which the other device is
connected; or consult Technical Support.
Caution:
To prevent internal heat buildup, allow sufficient air circulation around the console.
Do not place the console within .3 meters (1 foot) of any obstructive surface.
Caution:
Use and store the console in a location away from heat sources such as radiators or air
ducts, direct sunlight, excessive dust, mechanical vibration, or shock.
Caution:
Do not operate the equipment at altitudes higher than 3,000 meters (9,842 feet). Do
not store or operate the equipment in conditions outside of the stated equipment
storage, transport, and operating environmental conditions.
Caution:
Allow the console to acclimate for 24 hours before use after removal from storage.
Caution:
Risk of data loss. Do not put any magnetic media near or on devices that create any
magnetic fields.