Brevera Breast Biopsy System User Guide
Chapter 7: Biopsy
Page 70 MAN-04303-002 Revision 006
Device Driver
Warning:
Do not place fingers on the metal parts of the device driver. These parts can
move before a biopsy needle is installed.
1. Slide the device driver onto the device driver holder on the console until it clicks.
2. Confirm that the device driver cord is connected to the receptacle. The receptacle is
on the left side of the console, near the saline pinch valve.
Note
The console beeps three times before resetting the position of the metal parts of the
device driver.
3. Wait for the device driver to home before attaching the biopsy needle.
4. Press the Check button on the Technologist display if the device driver needs to be
rehomed.
Biopsy Needle and Tubing
Warning:
Before use, inspect the protective packaging and needle to verify that neither
has been damaged during shipment. If it appears that the packaging or
needle has been compromised, do not use the needle.
Warning:
Do not connect the biopsy needle until the system is powered on and the
device driver is homed and ready.
1. Place the sealed packaging for the biopsy needle on the console tray.
2. Peel open the sterile packaging of the biopsy needle.
3. Remove the biopsy needle from the tray, leaving the tubing and the tissue filter in the
tray. Make sure the biopsy needle gears are fully forward, toward the needle tip.
Warning:
To maintain sterility, leave the protective sheath on the tip of the biopsy
needle.