110: Set the sweeptime to 500 ms. Notice that upper/lower case in
commands does not matter. Also spaces before the suffix (“MS”)
are not required in SCPI.
120 and 130: Set markers 1 and 2 to a fixed value. Notice that the
value for marker 2 does not end with a frequency suffix. Hertz is a
default terminator and is understood.
140: Wait until the source has completed setting up the commands
that have been sent so far before turning on the output.
150: The ENTER statement causes the program to wait here until
the source responds to the previous *OPC? with a ‘1’.
160: The source has now completed processing the commands.
The RF frequency, power, and markers are at their programmed
values. Turn on the RF output of the source.
170: Select a continuously initiated sweep instead of the default
mode of non-continuous that was selected with *RST.
180: Clear the computer’s display.
190 to 220: Print a message on the computer’s display.