Module
Select
AUTO
Function Group
POWER and FREQUENCY
Menu Map
2 and 5
Description
This command sets the automatic selection of the millimeter source
module interface connector. The swept signal generator looks at both
front and rear connectors and determines the type of source module
(if any) connected. If a source module is present at both connectors,
the swept signal generator selects the front connector as the active
one. After selecting the interface the instrument frequency limits
and multiplier are altered accordingly. However, the leveling point is
not changed. See Leveling Point Module to set the swept signal
generator to level at the output of the source module.
An asterisk next to the key label indicates this feature is active. This
feature is the default after preset.
Programming Codes
SCPI:
SYSTem:MMHead:SELect:AUTO
ONjOFFlllO
SYSTem:MMHead:SELect:AUTO?
Analyzer: NONE
See Also
Module Menu
Module
Select
Front
Function Group
POWER and FREQUENCY
Menu Map
2 and 5
Description
This command causes the swept signal generator to examine only
the front panel source module interface connector to determine
the type of source module (if any) connected. The instrument
frequency limits and multiplier are altered according to the source
module connected. However, the leveling point is not changed. See
Leveling Point Module to set the swept signal generator to level
at the output of the source module.
An asterisk next to the key label indicates this feature is active.
M-24 Operating and Programming Reference